Phone

0822 5870 6420 (Anto)

0813 9929 1909 (Fikri)

Phone Office

+62-21 29563045





×


MX1102 Carbon Dioxide / Temp / RH Data Logger





Inquiry

Description

Onset’s MX1102 CO2 logger makes it more convenient than ever to measure and record CO2 in buildings and other non-condensing environments. It measures CO2 from 0 – 5,000 parts per million (ppm)


Feature

•Wireless communication via Bluetooth Low Energy (BLE) Technology
•Six month battery life at 5-minute CO2 logging rate
•Visual and audible high & low alarm thresholds
•Self-calibrating NDIR CO2 sensor technology ensures optimal accuracy and lower maintenance costs
•Patented connectivity technology


Spesification


Temperature Sensor

Range

0° to 50°C (32° to 122°F)

Accuracy

±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F)

Resolution

0.024°C at 25°C (0.04°F at 77°F)

Drift

<0.1°C (0.18°F) per year

 


RH Sensor

Range

1% to 70% RH when CO2 sensor is enabled (non-condensing)?
1% to 90% RH when CO2 sensor is disabled (non-condensing)

Accuracy

±2% from 20% to 80% typical at 25°C (77°F)

Hysteresis

±2% RH

Resolution

0.01% at 25°C (77°F)

Drift

<1% per year typical

 


CO2 Sensor

Range

0 to 5,000 ppm

Accuracy

±50 ppm ±5% of reading at 25°C (77°F), less than 70% RH and 1,013 mbar

Warm-up Time

15 seconds

Calibration

Auto or manual to 400 ppm

Non-linearity

<1% of FS

Pressure Dependence

0.13% of reading per mm Hg (corrected via user input for elevation/altitude)

Operating Pressure Range

950 to 1,050 mbar (use Altitude Compensation for outside of this range)

Compensated Pressure Range

-305 to 5,486 m (-1,000 to 18,000 ft)

Sensing Method

Non-dispersive infrared (NDIR) absorption

 


Response Time

Temperature

12 minutes to 90% in airflow of 1 m/s (2.2 mph)

RH

1 minute to 90% in airflow of 1 m/s (2.2 mph)

CO2

1 minute to 90% in airflow of 1 m/s (2.2 mph)

 


Logger

Radio Power

1 mW (0 dBm)

Transmission Range

Approximately 30.5 m (100 ft) line-of-sight

Wireless Data Standard

Bluetooth Smart (Bluetooth Low Energy, Bluetooth 4.0)

Logger Operating Range

0° to 50°C (32° to 122°F); 0 to 95% RH (non-condensing)

Logging Rate

1 second to 18 hours

Logging Modes

Fixed interval (normal, statistics) or burst

Memory Modes

Wrap when full or stop when full

Start Modes

Immediate, push button, date & time, or next interval

Stop Modes

When memory full, push button, date & time, or after a set logging period

Time Accuracy

± 1 minute per month at 25°C (77°F)

Power Source

4 AA 1.5 Volt batteries (user replaceable) or USB power source (5 V DC, 2 Watts)

Battery Life

6 months, typical with logging and sampling intervals of 5 minutes or slower; 6 months or less with logging and sampling intervals faster than 5 minutes while logging CO2. Entering burst logging mode will impact battery life. With HOBOmobile use, battery life can be reduced by remaining connected, excessive readouts, checking of Full Status Details, audible alarms, and paging. Visual/audible alarms and other events can have a marginal impact on battery life.

Memory

128 KB (84,650 measurements, maximum)

Download Type

USB 2.0 interface or via Bluetooth Smart

Full Memory Download Time

20 seconds via USB; approximately 60 seconds via Bluetooth Smart, may take longer the further the device is from the logger

LCD

LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range

Size

7.62 x 12.95 x 4.78 cm (3.0 x 5.1 x 1.88 inches)

Weight

267.4 g (9.43 oz)

Environmental Rating

IP50