SL500 Thermocouple Data Logger

SL500 Thermocouple Data Logger



Order

Description

Supco SL500 Series Data Logger for Measuring and Recording Temperature Using External Thermocouple Probe


Feature

  • Measure and Record Temperature Using Type J, K, E, N, T, S, R, B and C Thermocouple
  • Measuring -250°C to 1300°C Temp Range with ±0.7°C Accuracy (Included Thermocouple)
  • LCD Shows Current Temperature Readings, Battery Status and Alarm Status
  • Logs up to 32,638 Temperature Readings
  • 1 Second to 9 Hour User-Selectable Sampling Rate
  • User-Replaceable AAA Batteries with 2 Year Life (Dependant upon Environment and Sampling Rate)

Spesification

Temperature Thermocouple Types: J, K, E, N, T, S, R, B, C 
Resolution (Logged): 0.1°C (0.2°F)
Resolution (Display): 0.1°C (0.2°F) @ Up to 200°C
Cold Junction Accuracy: ±0.3°C (±0.5°F)
Supported Thermocouples Impedance: 50 Ohms Max 
K: -250°C to 1300°C (-418°F to 2372°F) @ ±0.7°C Accuracy
J: -170°C to 1000°C (-274°F to 1832°F) @ ±0.6°C Accuracy
E: -200°C to 820°C (-328°F to 1508°F) @ ±0.4°C Accuracy
N: -200°C to 1300°C (-328°F to 2372°F) @ ±0.8°C Accuracy
T: -200°C to 400°C (-328°F to 752°F) @ ±0.7°C Accuracy
S: -50°C to 1760°C (-58°F to 3200°F) @ ±2.8°C Accuracy
Type R: -50°C to 1760°C (-58°F to 3200°F) @ ±2.8°C Accuracy
Type B: 20°C to 1820°C (68°F to 3308°F) @ ±3.3°C Accuracy
Type C: 0°C to 2320°C (32°F to 4208°F) @ ±1.7°C Accuracy
Thermocouple
(Included)
Type: K - Insulated Beaded Wire
Length: 1 Meter (40")
Measurement Rating: -40°C to 260°C (-40°F to 500°F)
Data Storage Capacity 32,638 Temperature Readings
Sampling Rate 1 Second to 9 Hours
Starting Modes Programmable Delayed Start or Push to Start
Logging Modes Stop When Full or Rollover First In First Out (FIFO)
Power Source 3 AAA Batteries (Alkaline or Lithium)
Battery Life Typically 2 Years
Note: Dependant on Sampling Rate and Environment
Display Functions Current Reading and Time, Logging Status, Alarm Status, Recorded Sample History and Battery Life
Alarm Visual - Blinking LED in Display for High/Low Alarm Thresholds
Time Accuracy ±100 ppm @ 24°C (74°F)
OS Compatibility Windows XP, Vista, 7 (32-bit and 64-Bit Versions)
Calibration Single point offset calibration available through software
Standards Compliance CE
Dimensions 55mm x 57mm x 30mm (2.15" x 2.25" x 0.55")