Description
Catatan hingga 160 inci curah hujan dengan kecepatan sampai 12,7 cm (5 inci) per jam. Sistem Data Logging Rain Gauge bertenaga baterai dan dilengkapi dengan data log peristiwa HOBO® Pendant Event dengan alat pengukur hujan tip. Mudah mengumpulkan data curah hujan, waktu, dan durasi, serta suhu saat digunakan dengan perisai radiasi opsional (RS1 Solar Radiation Shield, atau M-RSA). Stasiun pangkalan atau antar jemput diperlukan.
Feature
- High quality tipping-bucket rain gauge includes a HOBO Pendant Event data logger. Base station or shuttle is required.
- Time and date stamp is stored for each tip - provides detail needed to determine rainfall rates and duration.
- Reliable, field-proven rain gauge.
- Compatible with HOBOware and HOBOware Pro software for logger setup, graphing and analysis.
- Also records temperature when used with an optional solar radiation shield.
Spesification
Rain Gauge
Maximum rainfall rate: 12.7 cm (5 in.) per hour
Calibration accuracy: ±1.0% (up to 1 in./hour for the RG3 or 20 mm/hour for the RG3-M)
Resolution: 0.01 in. (RG3) or 0.2 mm (RG3-M)
Calibration: Requires annual calibration: can be field calibrated or returned to the factory for re-calibration
Operating temperature range: 0° to +50°C (+32° to +122°F)
Storage temperature range: -20° to +70°C (-4° to +158°F)
Environmental rating: Weatherproof
Housing: 15.24cm (6-in.) aluminum bucket
Tipping bucket mechanism: Stainless steel shaft with brass bearings
Dimensions: 25.72 cm height x 15.24 cm diameter (10.125 x 6 in.); 15.39 cm (6.06 in.) receiving orifice
Weight: 1.2 Kg (2.5 lbs)
Part numbers: RG3 (0.01 in. per tip) ; RG3-M (0.2 mm per tip)
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
Logger
Time stamp: Resolution 1.0 second
Time accuracy: ± 1 minute per month at 25°C (77°F), see Plot B on page 10.
Operating range: -20° to 70°C (-4° to 158°F)
Environmental rating (for logger used outside of rain gauge): Tested to NEMA 6 and IP67; suitable for deployment outdoors
NIST traceable certification: Available for temperature only at additional charge; temperature range -20° to 70°C (-4° to 158°F)
Battery: CR-2032 3V lithium battery; 1 year typical use
Memory: 64K bytes – 16K to 23K tips when recording rain only; 25K to 30K data points when recording rain and temperature
Materials: Polypropylene case; stainless steel screws; Buna-N o-ring; PVC cable insulation
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
Temperature Measurement
Measurement range: -20° to 70°C (-4° to 158°F)
Accuracy: ± 0.47°C at 25°C (± 0.85°F at 77°F). A solar radiation shield is required for accurate temperature measurements in sunlight.
Resolution: 0.10°C at 25°C (0.18°F at 77°F), see Plot A on page 10.
Drift: Less than 0.1°C/year (0.2°F/year)
Response time: Airflow of 1 m/s (2.2 mph): 10 minutes, typical to 90%